Journal of Analytical Atomic Spectrometry · 1998 · 23 citations · 8 references
PharmacyChemistryAnalytical ChemistryElemental AnalysisMass Spectrometric AnalysisDetector SystemsMass AnalyzersMass Spectrometric Elemental AnalysisMass SpectrometryDetection Limit AssessmentInstrument Detection LimitMass Spectrometry CharacterizationMethod Detection LimitOxide Removal
A multi-strand Nafion® dryer, located between the spray chamber and torch of a plasma source mass spectrometer, removed about 97% of the total mass of water leaving the spray chamber. A significant reduction in the ratio of the oxide signal to that of the metal for cerium and barium was observed. The argon oxide and hydroxide signals at m/z 56 and 57, respectively, were also decreased by about a factor of 10 for m/z 56. In turn, an improvement in the detection limit for iron at this m/z by about a factor of 5 was obtained. The use of the dryer did not require re-optimization of the instrument operating conditions and, for the relatively clean solutions introduced by flow injection used in this study, did not cause any significant memory effect or peak broadening.
8
Jinfu Yang, Timothy S. Conver, John A. Koropchak et al. · Spectrochimica Acta Part B Atomic Spectroscopy · 1996 · 34 citations