Measurement Science and Technology · 2012 · 36 citations · 26 references
EngineeringRadiologyNuclear InstrumentationRadiation DetectionRadiation MeasurementDetector TechnologyRadiation DetectorsRadiation Imaging DetectorsX-ray DetectionX-ray DetectorsX-ray OpticsFilter DesignRadiation Flux
A band-pass flat-response x-ray detector is designed to measure the absolute M-band x-ray flux. The detector comprises an x-ray diode and a compound filter that is carefully designed to achieve the desired response function in the range from 1.6 to 4.4 keV, i.e. the flatness of the spectral response is better than 5%. The designed response function is in excellent agreement with the calibrated one, indicating that the x-ray detector with various responses can be achieved with the state-of-art fabrication technique.
26
S. Bastiani, A. Rousse, J. P. Geindre et al. · Physical review. E, Statistical physics, plasmas, fluids, and related interdisciplinary topics · 1997 · 166 citations
Natural Sciences, Physics, Optics +16
Plastic ablator ignition capsule design for the National Ignition Facility
D. S. Clark, S. W. Haan, B. A. Hammel et al. · Physics of Plasmas · 2010 · 97 citations · Full text