Biomedical Chromatography · 2009 · 118 citations · 36 references
PharmacyBiochemistryAnalytical ChemistryMass Spectrometric AnalysisSeparation ScienceAnalytical ChromatographyBioanalytical Method ValidationChromatographyInstrumental AnalysisLiquid ChromatographyMass SpectrometryAmino Acid AnalysisAmino Acid DerivatizationAmino Acid QuantificationAmino Acid Separation TechniquesAmino AcidsHigh-performance Liquid Chromatography-mass SpectrometryLiquid Chromatography-mass SpectrometryLiquid Chromatography-mass Spectrometry Analysis
An automated method for high-throughput amino acid analysis, using precolumn derivatization high-performance liquid chromatography/electrospray mass spectrometry (HPLC/ESI-MS), was developed and evaluated. The precolumn derivatization step was performed in the reaction port of a home-built auto-sampler system. Amino acids were derivatized with 3-aminopyridyl-N-hydroxysuccinimidyl carbamate, and a 3 microm Wakosil-II 3C8-100HG column (100 x 2.1 mm i.d.) was used for separation. To achieve a 13 min cycle for each sample, the derivatization and separation steps were performed in parallel. The results of the method evaluation, including the linearity, and the intra- and inter-precision, were sufficient to measure physiological amino acids in human plasma samples. The relative standard deviations of typical amino acids in actual human plasma samples were below 10%.
36
Fluorescence reaction for amino acids
Marc Roth · Analytical Chemistry · 1971 · 1.4K citations
Steven A. Cohen, Dennis P. Michaud · Analytical Biochemistry · 1993 · 978 citations