Clinical evaluation of a new hollow fibre membrane oxygenator

Filip De Somer, Dominique De Smet, M. Vanackere, Guido Van Nooten, Frank Caes, J. Delanghe

Perfusion · 1994 · 12 citations · 8 references

Concepts

Abstract

Twelve consecutive patients undergoing elective cardiac surgery were perfused with the Cobe Optima hollow fibre oxygenator. Gas transfer characteristics and blood handling were studied. The device had a maximum oxygen transfer of 315 ml/minute. The average shunt fraction was 4.5%, and was not influenced by blood-flow rate. Mean platelet count declined slightly to 91% of the baseline at the end of the study period. Haemolysis was evaluated by monitoring serum-free haemoglobin, serum haptoglobin and serum haemopexin. The evolution was as follows: free haemoglobin increased from 14 +/- 5 mg/100 ml to 85 +/- 0.8/100 ml (p = 0.01) at the end of bypass; haptoglobin decreased from 1.33 +/- 0.90 g/l to 0.89 +/- 0.15 g/l (p = 0.01); and haemopexin decreased from 0.84 +/- 0.13 g/l to 0.74 +/- 0.15 g/l (p = nonsignificant). In all patients the residual capacity of serum haptoglobin to protect against haemolysis was satisfactory. All patients had an uneventful postoperative course.