Publication | Closed Access
A comparative evaluation of the biostability of a poly (ether urethane) in the intraocular, intramuscular, and subcutaneous environments
15
Citations
20
References
1992
Year
Tissue EngineeringEngineeringSurgeryPrototype Peu LensesBiomedical EngineeringMinimum AccelerationVitreous BodyCataractPolymer ChemistryComparative EvaluationOphthalmologyTransparent PolyOcular TissueExperimental OphthalmologyPolymer ScienceSubcutaneous EnvironmentsWound HealingGlaucomaMedicinePlastic Surgery
A transparent poly (ether urethane) (PEU) was considered for use as a foldable intraocular lens material. The PEU was found to possess excellent mechanical, optical, and surface characteristics for this application. In vitro hydrolytic and ultraviolet aging studies suggested the PEU to be tolerant to conditions simulating 3-10 years of normal intraocular exposure. Different behavior was obtained, however, from intraocular and subcutaneous implantation of the PEU. After 6 months of intraocular exposure in the feline model, prototype PEU lenses had lost most or all of their optical resolving power. SEM analysis demonstrated scattered pitting and cracking on the lens surfaces. Degradation was found to be more extreme after as little as 30 days of subcutaneous exposure in rabbits. Severe pitting over the entire surface of implanted flat PEU specimens was observed by SEM. Macroscopic examination showed the samples to be frosty in appearance. It was postulated that the subcutaneous implant environment provides an accelerated in vivo model for materials intended for intraocular use. A minimum acceleration of 6-10x was estimated on a preliminary basis. The PEU studied here was found to be unsuitable for use as a foldable intraocular lens material.
| Year | Citations | |
|---|---|---|
Page 1
Page 1