Clinical Chemistry and Laboratory Medicine (CCLM) · 2005 · 35 citations · 7 references
Natural SciencesPharmaceutical AnalysisPharmacologyAnalytical ChemistryMass Spectrometric AnalysisMetabolomicsChromatographyInstrumental AnalysisLiquid ChromatographyMass SpectrometryMass Spectrometry-based MetabolomicsMetabolite QuantificationChromatography-based MetabolomicsInternal Standard AnalysisLiquid Chromatography-mass SpectrometryLiquid Chromatography-mass Spectrometry AnalysisLiquid Chromatography-mass Spectrometry Method DevelopmentMass Spectrometry-based Metabolite ProfilingTandem Mass SpectrometryTandem Mass Spectrometry Analysis
In a growing number of laboratories the technique of liquid chromatography-tandem mass spectrometry is used for the quantification of cyclosporin A in whole blood, employing cyclosporin D as the internal standard. Cyclosporin A is extensively metabolized in vivo; in liquid chromatography-tandem mass spectrometry respective metabolites can give rise to both parent and product ions that are isobaric with ions commonly used for the detection of cyclosporin A and cyclosporin D, respectively. In this article it is demonstrated that limited chromatography with co-elution of such metabolites together with cyclosporin A and cyclosporin D can lead to incorrect results.
7
Frank Streit, Victor W. Armstrong, Michael Oellerich · Clinical Chemistry · 2002 · 192 citations · Full text
Pharmacy, Immunology, Pharmacology +17
Therese Koal, Michael Deters, Bruno Casetta et al. · Journal of Chromatography B · 2004 · 127 citations