Analytical Chemistry · 1996 · 128 citations · 14 references
Natural SciencesChemistryMaterials ScienceBiosensorsPolymer ChemistrySensorsChemical SensorsFluorescence SpectroscopyLuminescent MaterialsOptical SensorsPolymer MembranesSensor MaterialsChemical Optical SensorsExcited State QuenchingLuminescent SensorsOxygen SensingOxygen SensorsPlasticizer EffectsPolymer MatricesPorphyrin Complexes
The design of luminescent oxygen sensors is guided by optimizing sensitivity and/or the form of the calibration function. Both qualities are governed by the molecular processes of luminescence quenching. To evaluate the influence of matrix effects, we prepared membranes based on oxygen-sensitive phosphorescent complexes of porphyrin ketones dissolved in plasticizer-free poly(vinyl chloride) (PVC) and polystyrene (PS). In a PVC matrix, both platinum(II) and palladium(II) octaethylporphyrin ketones exhibited perfectly linear Stern-Volmer intensity plots and almost single-exponential excited state decays. In a PS matrix, the sensitivity of palladium(II) octaethylporphyrin ketone was among the highest reported to date. Yet, slightly nonlinear Stern-Volmer plots and nonexponential decays illustrate the significance of matrix effects of PS. Addition of plasticizers to PVC-based sensors allowed tuning of the oxygen sensitivity in a wide range, while the Stern-Volmer plots became pronouncedly nonlinear. For the plasticizer bis(2-ethylhexyl) adipate, the decay profile was single-exponential in the absence but nonexponential in the presence of oxygen, which is attributed to a distribution of quenching rate constants.
14
Aleksander Siemiarczuk, Brian D. Wagner, William R. Ware · The Journal of Physical Chemistry · 1990 · 200 citations