The Accuracy of a Near-Infrared Spectroscopy Cerebral Oximetry Device and Its Potential Value for Estimating Jugular Venous Oxygen Saturation
74
Citations
40
References
2014
Year
The FORE-SIGHT cerebral oximeter was able to estimate oxygen saturation within the tissues of the frontal lobe under conditions of normocapnia and varying degrees of hypoxia (with 95% confidence interval of [-5.60 to 5.78] with ipsilateral blood sample data). These findings from healthy volunteers also suggest that the use of the calculated SnvO2 derived from SctO2 and SpO2 values may be a reasonable noninvasive method of estimating SjvO2 and therefore global cerebral oxygen consumption in the clinical setting. Further laboratory and clinical research is required to define the clinical utility of near-infrared spectroscopy determination of SctO2 and SnvO2 in the operating room setting.