A randomised study of peri-prosthetic bone density after cemented<i>versus</i>trabecular fixation of a polyethylene acetabular component
12
Citations
37
References
2011
Year
EngineeringPeri-prosthetic Bone DensityBone RepairRandomised StudySurgeryBiomedical EngineeringOsteoporosisOrthopaedic SurgeryBiomechanicsBone RemodelingOsteoarthritisPolyethylene Acetabular ComponentJoint ReplacementAcetabular BoneIdeal Acetabular ComponentBone ImagingDental BiomechanicsPolyethylene Articular SurfaceMedicine
The ideal acetabular component is characterised by reliable, long-term fixation with physiological loading of bone and a low rate of wear. Trabecular metal is a porous construct of tantalum which promotes bony ingrowth, has a modulus of elasticity similar to that of cancellous bone, and should be an excellent material for fixation. Between 2004 and 2006, 55 patients were randomised to receive either a cemented polyethylene or a monobloc trabecular metal acetabular component with a polyethylene articular surface. We measured the peri-prosthetic bone density around the acetabular components for up to two years using dual-energy x-ray absorptiometry. We found evidence that the cemented acetabular component loaded the acetabular bone centromedially whereas the trabecular metal monobloc loaded the lateral rim and behaved like a hemispherical rigid metal component with regard to loading of the acetabular bone. We suspect that this was due to the peripheral titanium rim used for the mechanism of insertion.