Concepedia

Publication | Closed Access

The effects of different steam‐sterilization programs on material properties of poly(<scp>L</scp>‐lactide)

56

Citations

18

References

1991

Year

Abstract

As-polymerized poly(L-lactide) test rods were sterilized by seven different specially designed computer-operated autoclaving programs. As a control, common hospital sterilization was performed. In all cases, the molecular weight decreased after sterilization. A short time high-temperature sterilization led to less molecular weight decrease than a low sterilization temperature cycle with a longer sterilization time. Regular hospital sterilization significantly reduced the elongation at break and also resulted in a decrease of 35% in tensile strength. The program causing minimal damage to the material properties was studied in detail. This program, with a sterilization period of 60 s and 129 degrees C, was effective for PLLA sterilization and also looks very promising for sterilization of other thermo- and moisture-labile polymers.

References

YearCitations

Page 1