Hot Wire Anemometer in Liquid Mercury

Miklos Sajben

Review of Scientific Instruments · 1965 · 38 citations · 0 references

Concepts

Abstract

A hot wire system has been developed which is suitable for operation in liquid mercury at speeds from 1 to 12 cm/sec. The wire was 38 μ diam tungsten with 2.5 μ thick enamel coating, used with 2 A heating current in constant current operation. The Nusselt number was determined experimentally as a function of Péclet number in the range 0<Pé<1.2, and was found to be compatible with available theory. Fluid impurities posed a serious problem which was countered with a special calibration procedure. The associated bridge circuit is described. Manipulation of bridge signal is made difficult by the low frequencies associated with low speeds, but can be handled by available techniques.