Rapid and accurate direct antibiotic susceptibility testing of blood culture broths using MALDI Sepsityper combined with the BD Phoenix automated system

Briony Hazelton, Lee Thomas, Thomas Olma, Jen Kok, Matthew O’Sullivan, Sharon C.‐A. Chen, Jonathan R. Iredell

Journal of Medical Microbiology · 2014 · 27 citations · 21 references

Concepts

Abstract

Antibiotic susceptibility testing with the BD Phoenix system on bacterial cell pellets generated from blood culture broths using the Bruker MALDI Sepsityper kit was evaluated. Seventy-six Gram-negative isolates, including 12 with defined multi-resistant phenotypes, had antibiotic susceptibility testing (AST) performed by Phoenix on the cell pellet in parallel with conventional methods. In total, 1414/1444 (97.9 %) of susceptibility tests were concordant, with only 1 (0.07 %) very major error. This novel method has the potential to reduce the turnaround time for AST results by up to a day for Gram-negative bacteraemias.

References