Two-Detector, 512-Element High Purity Germanium Camera Prototype
18
Citations
6
References
1978
Year
EngineeringRadiation DetectionPhysicsGamma-ray CameraCalibrationHealth SciencesScintillatorApplied PhysicsIntrinsic Spatial ResolutionShallow Orthogonal GroovesDetector PhysicInstrumentationRadiation ImagingDetector PhysicsImage SensorRadiologyX-ray Imaging
A gamma-ray camera consisting of two 3.2×3.2×1-cm3 HPGe detectors has been assembled. Shallow orthogonal grooves define 512 2×2-mm2 elements. Square hole collimators have been fabricated with design parameters that exploit the unique characteristics of the detector. Intrinsic spatial resolution is a square function with 2-mm width, and energy resolution is approximately 2.5% FWHM at 140 keV. Evidently superior images are obtained when this instrument is compared to state-of-the-art scintillation cameras.