European Journal of Mass Spectrometry · 2000 · 108 citations · 11 references
Natural SciencesChemistryAnalytical ChemistryInstrumentationMass Spectrometric AnalysisAccurate Mass MeasurementDetector SystemsInstrumental AnalysisMass AnalyzersMass SpectrometryElectrospray IonizationFull Width At Half MaximumHigh-resolution Mass AnalysisMass Accuracy EvaluationMeasurement AccuracySpectrometer DesignTime-of-flight Mass Spectrometry
A concept of an “ideal” orthogonal-injection time-of-flight (O-TOF) mass spectrometer is proposed and a new high-resolution O-TOF mass spectrometer with an electrospray ion source and a molecule-ion reactor is described. The instrument design was optimized for high resolution and sensitivity using the concept of an ideal O-TOF mass spectrometer. A mass resolving power of about 20,000 (full width at half maximum) and a mass measurement accuracy in the 10 ppm range is demonstrated.
11
Howard R. Morris, Thanai Paxton, Anne Dell et al. · Rapid Communications in Mass Spectrometry · 1996 · 434 citations
Andrew N. Krutchinsky, Vic Spicer, Ragnar G. Dworschak et al. · Rapid Communications in Mass Spectrometry · 1998 · 137 citations