Natural SciencesPhysicsRadiologyDiagnostic ImagingDiagnostic RadiologyNuclear PhysicsGamma-ray Data AnalysisGamma-ray DetectorsGamma-ray InstrumentationRadiation Imaging DetectorsRadiation Imaging SystemsRadiographic ImagingBackprojectionDetector Spatial ResolutionGamma Camera ImagingGamma-ray ImagingGamma-ray TomographyHigh-purity Germanium DetectorsImaging ResolutionInteraction Point Mapping
The use of a high purity germanium gamma camera capable of assigning the interaction point of a gamma-ray unambiguously to a 2×2mm2 element, together with the use of matched square-hole tantalum tube collimators, permits the application of a weighted backprojection technique to reconstruct images that do not show the loss of spatial resolution associated with distance between the object and collimator surface. The technique converges, and images with improved definition are obtained.