Publication | Closed Access
Corneal Thickness Measurements Using Time-domain Anterior Segment OCT, Ultrasound, and Scheimpflug Tomographer Pachymetry Before and After Corneal Cross-linking for Keratoconus
37
Citations
29
References
2012
Year
Visante pachymetry shows better agreement with USP compared to Pentacam after CXL, which may be due to the inhomogeneous reflectivity of the postoperative cross-linked cornea and possibly altered refractive index and acoustic impedance that may influence the observed differences among techniques.