Parameter evaluation and fully-automated radiosynthesis of [ 11 C]harmine for imaging of MAO-A for clinical trials

C. Philippe, Markus Zeilinger, Markus Mitterhauser, Monika Dumanic, Rupert Lanzenberger, Marcus Hacker, Wolfgang Wadsak

Applied Radiation and Isotopes · 2015 · 18 citations · 12 references

DOIFull text

Open access

Concepts

Abstract

The aim of the present study was the evaluation and automation of the radiosynthesis of [(11)C]harmine for clinical trials. The following parameters have been investigated: amount of base, precursor concentration, solvent, reaction temperature and time. The optimum reaction conditions were determined to be 2-3mg/mL precursor activated with 1eq. 5M NaOH in DMSO, 80°C reaction temperature and 2min reaction time. Under these conditions 6.1±1GBq (51.0±11% based on [(11)C]CH3I, corrected for decay) of [(11)C]harmine (n=72) were obtained. The specific activity was 101.32±28.2GBq/µmol (at EOS). All quality control parameters were in accordance with the standards for parenteral human application. Due to its reliability and high yields, this fully-automated synthesis method can be used as routine set-up.

References

12