Concepedia

Publication | Closed Access

Factors influencing the accuracy of oscillometric blood pressure measurement in critically ill patients

125

Citations

20

References

2003

Year

Abstract

The new algorithm reduced the overall bias of the oscillometric method but still showed a significant discrepancy between both methods of blood pressure measurement, primarily due to the mismatch between upper arm circumference and cuff size. The improvement of the algorithm alone could not result in a sufficient improvement of oscillometric blood pressure measurement. A reevaluation of the recommendations concerning the relation between upper arm circumference and cuff size are urgently required if oscillometric blood pressure measurement should become a reasonable alternative to intra-arterial blood pressure measurement in critically ill patients.

References

YearCitations