Factors influencing the accuracy of oscillometric blood pressure measurement in critically ill patients
125
Citations
20
References
2003
Year
The new algorithm reduced the overall bias of the oscillometric method but still showed a significant discrepancy between both methods of blood pressure measurement, primarily due to the mismatch between upper arm circumference and cuff size. The improvement of the algorithm alone could not result in a sufficient improvement of oscillometric blood pressure measurement. A reevaluation of the recommendations concerning the relation between upper arm circumference and cuff size are urgently required if oscillometric blood pressure measurement should become a reasonable alternative to intra-arterial blood pressure measurement in critically ill patients.