Concepedia

Publication | Closed Access

Influence of operator experience on the performance of ultrasound biometry compared to optical biometry before cataract surgery

138

Citations

12

References

2003

Year

Abstract

Experienced US operators had less difference and lower variability in the difference between applanation US and IOLMaster readings for axial length and ACD measurements. The noncontact optical method, which is essentially operator independent, gave significantly more reliable biometry before cataract surgery, especially in the case of less experienced operators.

References

YearCitations