Catheter-Induced Intra-Hisian and Intrafascicular Block During Recording of His Bundle Electrograms
35
Citations
10
References
1974
Year
Heart FailureCardiac AnaesthesiaDevice TherapyMechanical InjurySurgeryElectrophysiological EvaluationRecording CatheterRight Bundle BranchCardiologyCardiac MechanicIntrafascicular BlockBundle ElectrogramsCardiac ArrestPatient SafetyElectromyographyElectrophysiologyAnesthesiaMedicineEmergency MedicineAnesthesiology
Mechanical injury of the atrioventricular (A-V) node, His bundle, and right bundle branch is a potential complication of His bundle recording. In the presence of left bundle branch block, catheter-induced injury of the right bundle branch may cause prolongation of A-V conduction time or even complete A-V block. Two cases are reported in which catheter-induced transient high grade Mobitz II A-V block, acute reversible prolongation of infranodal conduction time with probable intra-Hisian conduction delay, and persistent (6 hours duration) complete A-V block occurred in patients with left bundle branch block. The observations made indicate that A-V conduction times and the appearance of "split His" potentials may be influenced by the recording catheter itself, and suggest that the operators be prepared to institute immediate right ventricular pacing when His bundle recordings are obtained in patients with left bundle branch block.