Review of Scientific Instruments · 2014 · 77 citations · 5 references
Natural SciencesNuclear EngineeringPhysicsNuclear InstrumentationPlasma PhysicsRadiation DetectionCalibrationEquipment CalibrationHigh-temperature PlasmasPlasma DiagnosticsRadiometric CalibrationTokamak EngineeringElectron Cyclotron Emission DiagnosticsField Instrument CalibrationMagnetic DiagnosticsRadiometer CalibrationRelative CalibrationTokamak PhysicsTokamak PlasmasToroidal Magnetic Fields
A novel 32-channel electron cyclotron emission radiometer has been designed and tested for the measurement of electron temperature profiles on the HL-2A tokamak. This system is based on the intermediate frequency filter detection technique, and has the features of wide working frequency range and high spatial resolution. Two relative calibration methods have been investigated: sweeping the toroidal magnetic field and hopping the output frequency of the local oscillator. Preliminary results show that both methods can ensure reasonable profiles.
5
Preliminary results of ELMy H-mode experiments on the HL-2A tokamak
X.R. Duan, J.Q. Dong, L.W. Yan et al. · Nuclear Fusion · 2010 · 60 citations
The development of Thomson scattering system on HL-2A tokamak
Yuan Huang, P. Zhang, C. H. Liu et al. · Review of Scientific Instruments · 2007 · 32 citations