Physics in Medicine and Biology · 2005 · 78 citations · 12 references
MedicineOncologyRadiation OncologyClinical Radiation OncologyRadiation DosimetryTreatment PlanningExternal Beam RadiotherapyIntensity-modulated RadiotherapyRadiation TherapyRadiation Therapy For Prostate DiseaseRadiation Treatment PlanningTreatment Planning Dose CalculationBeam Delivery SystemsExternal Beam Radiotherapy DosimetryIntensity Modulated Radiation Therapy Dose CalculationIntensity-modulated Radiation Therapy DosimetryProstate Intensity-modulated RadiotherapySimulated AnnealingSimulated Annealing OptimizationTreatment Plan Optimization
Sweeping-window arc therapy (SWAT) is a variation of intensity-modulated radiation therapy (IMRT) with direct aperture optimization (DAO) that is initialized with a leaf sequence of sweeping windows that move back and forth periodically across the target as the gantry rotates. This initial sequence induces modulation in the dose and is assumed to be near enough to a minimum to allow successful optimization, done with simulated annealing, without requiring excessive leaf speeds. Optimal beam weights are calculated analytically, with easy extension to allow for variable beam weights. In this paper SWAT is tested on a phantom model and clinical prostate case. For the phantom, constant and variable beam weights are used. Although further work (in particular, improving the dose model) is required, the results show SWAT to be a feasible approach to generating deliverable dynamic arc treatments that are optimized.
12
Optimization by Simulated Annealing
Scott Kirkpatrick, C. D. Gelatt, M.P. Vecchi · Science · 1983 · 44.6K citations
Direct aperture optimization: A turnkey solution for step‐and‐shoot IMRT
David Shepard, Matthew A. Earl, X. A. Li et al. · Medical Physics · 2002 · 366 citations
Leaf sequencing algorithms for segmented multileaf collimation
Srijit Kamath, Sartaj Sahni, Jonathan Li et al. · Physics in Medicine and Biology · 2003 · 176 citations