Analytical Chemistry · 1994 · 94 citations · 22 references
PharmacyPharmacologyToxicologyAnalytical ChemistryAnalytical ToxicologyMass Spectrometric AnalysisAssay ValidationBioanalytical Method ValidationDrug DispositionInstrumental AnalysisMass SpectrometryTime-of-flight Mass Spectrometry
Analytical methods have been developed for the detection and quantification of cyclosporin A (CsA) in blood using time-of-flight secondary-ion mass spectrometry (TOF-SIMS) and matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI/TOF-MS). Linear calibration curves were obtained for both methods in the range 25-2000 ng/mL in whole blood. The limit of detection (LOD) and limit of quantification (LOQ) were determined to be 7 and 23 ng/mL, respectively, for the TOF-SIMS method; the LOD and LOQ were 10 and 33 ng/mL, respectively, for the MALDI/TOF-MS method. Coefficients of variation ranged from 3 to 5% for the TOF-SIMS method and from 4 to 8% for the MALDI/TOF-MS method. Blood samples were also analyzed by a nonspecific method, fluorescence polarization immunoassay (FPIA), and by a specific method, high-performance liquid chromatography (HPLC). The TOF-MS results are in good agreement with the HPLC results, but not with the FPIA results. The TOF-MS methods can also provide information about CsA metabolites.
22
Biological effects of cyclosporin A: A new antilymphocytic agent
J. F. Borel, Camille Feurer, H. U. Gubler et al. · Inflammation Research · 1976 · 1.5K citations
Thomas R. Covey, Eric C. Huang, Jack D. Henion · Analytical Chemistry · 1991 · 444 citations