Disposable electrodes for field screening of trace uranium

Joseph Wang, Rossi Setiadji

Electroanalysis · 1994 · 20 citations · 15 references

Concepts

Abstract

Abstract A novel electrochemical sensor for trace (μg/L) Uranium, based on adsorptive stripping potentiometry, is described. The sensor operation relies on the adsorptive accumulation of the uranium‐cupferron complex onto a mercury‐coated screen‐printed carbon electrode. The disposable nature of these electrodes makes them very attractive for on‐site characterization of nuclear energy sites. Various variables affecting the potentiometric stripping response were characterized and optimized to yield low‐detection limit (e.g., 0.5 μg/L following a 5 minute adsorption) and good precision (relative standard deviation (RSD) = 1.4% for n = 15, 25 μg/L uranium). Convenient measurements can be carried out in nondeaerated and quiescent solutions. Applicability to the quantitation of uranium in groundwater samples is demonstrated.