Electroanalysis · 1994 · 20 citations · 15 references
Natural SciencesChemical EngineeringChemistryAnalytical ChemistryElectrochemistrySensorsChemical SensorsElectrochemical DetectionElectrochemical SensorsEnvironmental Analytical ChemistryInorganic Elemental AnalysisTrace Element AnalysisCarbon-based Electrode MaterialsCarbon-based Working ElectrodesDisposable ElectrodesPotentiometric DetectionPotentiometric MeasurementPrecisionUranium AnalysisUranium Measurement
Abstract A novel electrochemical sensor for trace (μg/L) Uranium, based on adsorptive stripping potentiometry, is described. The sensor operation relies on the adsorptive accumulation of the uranium‐cupferron complex onto a mercury‐coated screen‐printed carbon electrode. The disposable nature of these electrodes makes them very attractive for on‐site characterization of nuclear energy sites. Various variables affecting the potentiometric stripping response were characterized and optimized to yield low‐detection limit (e.g., 0.5 μg/L following a 5 minute adsorption) and good precision (relative standard deviation (RSD) = 1.4% for n = 15, 25 μg/L uranium). Convenient measurements can be carried out in nondeaerated and quiescent solutions. Applicability to the quantitation of uranium in groundwater samples is demonstrated.
15
Adsorptive stripping voltammetry in trace analysis
R. Kalvoda, M. Kopanica · Pure and Applied Chemistry · 1989 · 180 citations · Full text