500-nm-Resolution 10 keV X-Ray Imaging Transmission Microscope with Tantalum Phase Zone Plates

Yasushi Kagoshima, Takashi Ibuki, Kengo Takai, Yoshiyuki Yokoyama, Naoki Miyamoto, Yoshiyuki Tsusaka, Junji Matsui

Japanese Journal of Applied Physics · 2000 · 14 citations · 12 references

Concepts

Abstract

An imaging transmission hard X-ray microscope has been constructed at the Hyogo-BL (BL24XU) of SPring-8. It makes use of X-ray phase zone plates (PZP's) made of tantalum as its condenser and objective lenses. The objective PZP has an outermost zone width of 250 nm, which corresponds to the theoretically expected spatial resolution of 300 nm. An experiment was performed at the photon energy of 10 keV to check the performance of the microscope. Since a 250 nm line-and-space pattern was clearly resolved, we concluded that the microscope attained a spatial resolution limit better than 500 nm. A few samples were also examined and the feasibility of the microscope was successfully demonstrated.