Concepedia

Publication | Open Access

Expiratory time constant for determinations of plateau pressure, respiratory system compliance, and total resistance

70

Citations

13

References

2013

Year

Abstract

We found the ƬE method was just as good as the EIP method for determining Pplt, Crs, and Rtot for various modes of ventilatory support for patients with acute respiratory failure. It is unclear if the ƬE method can be generalized to patients with chronic obstructive lung disease. ƬE is determined during passive deflation of the lungs without the need for changing the ventilatory mode and disrupting a patient's breathing. The ƬE method obviates the need to apply an EIP, allows for continuous and automatic surveillance of inspiratory Pplt so it can be maintained ≤ 30 cm H₂O for lung protection and patient safety, and permits real time assessments of pulmonary mechanics.

References

YearCitations