Physics in Medicine and Biology · 2006 · 95 citations · 6 references
EngineeringElectrical EngineeringMechatronics EngineeringElectromagneticsEquipmentInstrumentationCalibrationElectromagnetic CompatibilityTracking SystemsElectromagnetic TrackingElectromagnetic Tracking SystemsJitterJitter EstimationOffset Performance MonitoringPerformance MonitoringPositional Error AnalysisPrecision Positioning
This note uses a published protocol to evaluate a newly released 6 degrees of freedom electromagnetic tracking system (Aurora, Northern Digital Inc.). A practice for performance monitoring over time is also proposed. The protocol uses a machined base plate to measure relative error in position and orientation as well as the influence of metallic objects in the operating volume. Positional jitter (E(RMS)) was found to be 0.17 mm +/- 0.19 mm. A relative positional error of 0.25 mm +/- 0.22 mm at 50 mm offsets and 0.97 mm +/- 1.01 mm at 300 mm offsets was found. The mean of the relative rotation error was found to be 0.20 degrees +/- 0.14 degrees with respect to the axial and 0.91 degrees +/- 0.68 degrees for the longitudinal rotation. The most significant distortion caused by metallic objects is caused by 400-series stainless steel. A 9.4 mm maximum error occurred when the rod was closest to the emitter, 10 mm away. The improvement compared to older generations of the Aurora with respect to accuracy is substantial.
6
Accuracy assessment protocols for electromagnetic tracking systems
D.D. Frantz, Andrew D. Wiles, Stephen Leis et al. · Physics in Medicine and Biology · 2003 · 174 citations
Systematic distortions in magnetic position digitizers
Wolfgang Birkfellner, Franz Watzinger, Felix Wanschitz et al. · Medical Physics · 1998 · 128 citations · Full text
Stability of miniature electromagnetic tracking systems
Kurt Schicho, Michael Figl, Markus Donat et al. · Physics in Medicine and Biology · 2005 · 91 citations