Dentomaxillofacial Radiology · 1993 · 56 citations · 4 references
DentistryOral RadiologyDiagnostic ImagingDiagnostic RadiologyDigital ImagingDental ImagingDental RadiographyDigital Dental ImagingIntraoral RadiographyIntraoral ScanningRadiographic InterpretationDental Digital RadiographyDigital Intraoral RadiographyFilm ComparisonsFilm-based Intraoral RadiographyFilm-based RadiographyImage Quality BenchmarkingRadiographic Image QualityRadiographic Quality Assessment
The Sens-A-Ray digital imaging system was compared with group D- and E-speed intraoral radiographic film with respect to resolution, speed and image quality. Using a line-pair plate, resolution was determined to be 10 line-pairs per millimetre. With a constant potential X-ray source, the speed of the system was over three times that of E-speed emulsion and the response of pixel value to exposure was linear. Absorption curves with aluminium showed a similar response to D-speed film. An aluminium foil test object was used to compare the Sens-A-Ray system with both film types. Thirty observers were scored for successful identification of a pattern of holes in the test object. There was little difference between D- and E-speed film groups, and successful identification was significantly reduced with the Sens-A-Ray images when compared with D- and E-speed films.
4
Presentation and physical evaluation of RadioVisioGraphy
Francis Mouyen, Christoph Benz, E Sonnabend et al. · Oral Surgery Oral Medicine Oral Pathology · 1989 · 234 citations
Quantitative assessment of a new dental imaging system
K. A. Homer, J Czajka, A.C. Shearer et al. · British Journal of Radiology · 1991 · 30 citations