Improvements in the precision of in vivo bone lead measurements
38
Citations
6
References
1989
Year
EngineeringK-shell X-raysBiomedical EngineeringOsteoporosisOrthopaedic SurgeryX-ray FluorescenceX-ray ImagingBiomechanicsPb KInstrumentationRadiation ImagingRadiation OncologyNuclear MedicineRadiologyHealth SciencesMechanobiologyBone HealthMedical ImagingBone DensityRadiographic ImagingBone MetabolismBone Imaging
The measurement of Pb in bone by X-ray fluorescence using 109Cd gamma -ray to excite K-shell X-rays is considered. Improvements to the basic system performance were made by changing the design of the source collimator and by including data from the Pb K beta 1,3 photopeaks, having used pile-up rejection to reduce background in their region of the spectrum. The summed variances for eight sample measurements are given: two each of four plaster-of-Paris phantoms embedded in wax.