Audiology and Neurotology · 2011 · 35 citations · 14 references
Health ProfessionsOtolaryngologyCochlear ImplantsHearing ScienceOtologyDevice DesignEar AnatomyEar SurgeryImplant TechnologyMiddle Ear FunctionSurgical ImplantationAmplitude Ratio AnalysisMiddle Ear CouplingMiddle Ear ImplantsMiddle Ear PhysiologyRound Window MechanicsRound Window SurgeryStapes BiomechanicsStapes Motion
Round-window stimulation is a new clinical approach for the application of active middle-ear implants. To investigate factors influencing the efficiency of round-window stimulation, experiments in 6 human temporal bones were performed with different actuator geometries and coupling conditions. The experiments show that the amplitude ratio between stapes and round-window actuator vibration is most efficient when using a 1.0-mm diameter rod with a 30° inclined tip geometry and an attached silicone pad. In this case, the amplitude ratio is 0.34 for frequencies up to 1.5 kHz and 0.27 for frequencies up to 20 kHz, with a standard deviation of only 4-6 dB at most frequencies. The analysis of data presented here and in a companion paper suggests that control of proper round-window membrane pretension as well as the inclined tip geometry are the major requirements for maximal performance.
14
Application of the vibrant soundbridge® to unilateral osseous atresia cases
H Frenzel, Frauke Hanke, Millo Achille Beltrame et al. · The Laryngoscope · 2008 · 120 citations
Medicine, Audiology, Otolaryngology +17
Hideko Heidi Nakajima, Wei Dong, Elizabeth S. Olson et al. · Otology & Neurotology · 2010 · 101 citations · Full text