14 - pixel , multiplexed array of gamma-ray microcalorimeters with 47eV energy resolution at 103keV

W. B. Doriese, Joel N. Ullom, James A. Beall, William Duncan, L.F. Vieira Ferreira, G. C. Hilton, Robert D. Horansky, K. D. Irwin, John A. B. Mates, C. D. Reintsema,

Applied Physics Letters · 2007 · 70 citations · 11 references

Concepts

Abstract

The authors present a prototype for a high-energy-resolution, high-count-rate, gamma-ray spectrometer intended for nuclear forensics and international nuclear safeguards. The prototype spectrometer is an array of 14 transition-edge-sensor microcalorimeters with an average energy resolution of 47eV (full width at half maximum) at 103keV. The resolution of the best pixel is 25eV. A cryogenic, time-division multiplexer reads out the array. Several important topics related to microcalorimeter arrays are discussed, including cross-talk, the uniformity of detector bias conditions, fabrication of the arrays, and the multiplexed readout. The measurements and calculations demonstrate that a kilopixel array of high-resolution microcalorimeters is feasible.