Publication | Closed Access
Kohlrabi-Based Amperometric Biosensor for Hydrogen Peroxide Measurement
36
Citations
6
References
1991
Year
EngineeringKohlrabi-based Amperometric BiosensorChemistryChemical EngineeringOrganic ElectrochemistryBioanalysisCarbon Paste ElectrodeAnalytical ChemistryChemical SensorElectrode Reaction MechanismAbstract Hydrogen PeroxideBiochemistryHydrogenPossible InterferencesElectrochemistryOxygen Reduction ReactionNatural SciencesBioelectronicsElectroanalytical Sensor
Abstract Hydrogen peroxide was determined amperometrically in a steady-state arrangement by utilizing a kohlrabi - ferrocene based carbon paste electrode. A very short response time (2.6 seconds) and a relatively large usable pH range (5.0–7.4) were obtained. Several important hydrogen donors were studied as possible interferences.