Applied Optics · 2004 · 171 citations · 14 references
EngineeringBioengineeringBiomedical EngineeringBiomedical ImagingBiomedical OpticsBiophotonicsOptical BiopsyOptical ImagingPolarization OpticsPolarization-based ImagingTissue Optical PropertiesTissue OpticsAccuracyInstrument Calibration For Polarimetric ImagingMueller Matrix ImagingOptical PropertiesPolarimetric CalibrationPolarimetric ImagingPolarimetryPolarization Imaging
We present a new polarimetric imaging system based on liquid-crystal modulators, a spectrally filtered white-light source, and a CCD camera. The whole Mueller matrix image of the sample is measured in approximately 5 s in the transmission mode. The instrument design, together with an original and easy-to-operate calibration procedure, provides high accuracy over a wide spectral range (500-700 nm). This accuracy has been assessed by measurement of a linear polarizer at different orientations and a thick wedged quartz plate as an example of a partially depolarized retarder. Polarimetric images of a stained hepatic biopsy with significant fibrosis have been taken at several wavelengths. The optical properties of Picrosirius Red stained collagen (diattenuation, retardance, and polarizance) have been measured independently from each other between 500 and 700 nm.