Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · 2008 · 11 citations · 7 references
Transverse ResolutionsTissue ImagingRadiologyEngineeringMedical ImagingOphthalmologyMeasurementTransverse ResolutionHigh ResolutionBiomedical ImagingBiomedical OpticCoherenceOptical Coherence TomographyThree-dimensional Point-spread FunctionOptical ImagingTomography3D ImagingHealth Sciences
Two significant figures of merit for optical coherence tomography (OCT) systems are the axial and transverse resolutions. Transverse resolution has been defined using the Rayleigh Criterion or from Gaussian beam optics. The axial resolution is generally defined in terms of the coherence length of a Gaussian shaped source. Whilst these definitions provide a useful mathematical reference they are somewhat abstracted from the three dimensional resolution that is encountered under practical imaging conditions. Therefore, we have developed a three-dimensional resolution target and measurement methodology that can be used to calibrate the three-dimensional resolution of OCT systems.
7
David Huang, Eric A. Swanson, Charles P. Lin et al. · Science · 1991 · 13.5K citations
Ultrahigh-resolution ophthalmic optical coherence tomography
Wolfgang Drexler, Uwe Morgner, Ravi K. Ghanta et al. · Nature Medicine · 2001 · 949 citations · Full text
In vivo ultrahigh-resolution optical coherence tomography
Wolfgang Drexler, Uwe Morgner, Franz X. Kärtner et al. · Optics Letters · 1999 · 926 citations
Engineering, Microscopy, Coherence +16