Factors Influencing Long-Term in vivo Reproducibility of QCT (Vertebral Densitometry)
11
Citations
0
References
1993
Year
Essentially, the calibration procedure appeared to fail in some examinations of the problem cases, causing poor long-term reproducibility. This calibration failure may be related to positioning of the subject in the gantry and to variations in electrical parameters of the X-ray tube. A CV > 1% between the four slopes of calibration for the individual vertebrae measured at the same visit may indicate calibration problems and may suggest the need for repeat scanning.