Correlation of Solution and Gas Phase Complexation Assessed by Electrospray Ionization Mass Spectrometry:  Application to One-, Two-, and Three-Ring Macrocycles

Keshi Wang, George W. Gokel

The Journal of Organic Chemistry · 1996 · 63 citations · 32 references

Concepts

Abstract

Electrospray ionization mass spectrometry (ESI-MS) was used to probe multiple cation complexation by C(12)H(25)<N18N>(CH(2))(12)<N18N>(CH(2))(12)<N18N>C(12)H(25), 2, and <18N>CH(2)C(6)H(4)CH(2)<N18>, 3. Complexation of two cations (2Na(+), 2 K(+), or Na(+) and K(+)) by 3 and three cations by 2 (3 Na(+), 3 K(+), and mixtures) as well as mixed proton-metallic cation complexes of both were observed. The K(+)/Na(+) cation-binding selectivity of 18-crown-6 was studied by ESI-MS of a methanol solution, and the selectivity profile was favorably compared with data obtained previously by ion-selective electrode techniques in the same solvent.