Concepedia

Publication | Closed Access

Development of a low noise preamplifier for the detection and position determination of single electrons in a Cerenkov ring imaging detector by charge division

14

Citations

6

References

1988

Year

Abstract

A preamplifier using a low-noise dual-gate MOSFET front end has been designed, built, and tested. It performs well, having a noise level of about 500 electrons RMS at a shaping time of 65 ns. It is linear over its entire range to better than 1%. It has been used to detect single photoelectrons in a Cerenkov ring imaging detector. A single channel contains preamp, RC-CR shaper, gain adjustment, driver, and calibration circuitry. The circuit is described in detail, and results of noise and linearity measurements are presented.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">&gt;</ETX>

References

YearCitations

Page 1