An Acoustic Impedance Meter for Rapid Field Measurements

R. H. Bolt, A. A. Petrauskas

The Journal of the Acoustical Society of America · 1943 · 17 citations · 0 references

Concepts

Abstract

A method is developed for rapid, direct measurement of acoustic impedance. It is applicable to large panels and to materials already installed for use, and therefore offers a means of investigating effects of panel vibration and mounting conditions. The pressure and pressure gradient are measured by two microphones near the surface, in a line normal to it. If p1 and p2 are the sound pressures (time phase vectors) at the microphones; d their spacing; λ the wave-length; and θ the phase angle of the impedance; then the complex acoustic impedance ratio is given by zρc≃πdλ|p1+p2p1−p2|ei(θ−π/2). This approximation is reasonably good for λ/d greater than 10 or 15, depending somewhat on the impedance value. The microphone output voltages e1 and e2 are fed into an electronic circuit which gives voltages proportional to (e1+e2) and (e1−e2). These combination voltages are then mixed, and adjusted in relative phase and magnitude to give a null indication. The impedance magnitude and phase angle are determined from the dial settings. Null methods are also used for calibration and alignment, which must be performed with great precision to obtain reasonably accurate results. Preliminary tests with the instrument, using commercial dynamic microphones, appear to yield satisfactory results for frequencies from 50 to 500 c.p.s. and for impedance magnitudes up to 10 ρc.