Efficient Class-E design for inductive powering wireless biotelemetry applications

Saad Mutashar Abbas, M. A. Hannan, A. S. Salina

2012 · 25 citations · 7 references

Concepts

Abstract

This paper introduces a modified an efficient design for different Class-E power amplifiers suitable to be use in biomedical implantable devices and in biomedical wireless telemetry devices which is usually have a small load resistor (several Ohms) due to its inductive powering. The different Class-E amplifiers operated according to the frequency chart for the medical implant communications service (MICS) 27 MH <inf xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">Z</inf> , and industrial scientific and medical (ISM) (13.56 MH <inf xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">Z</inf> ), and low frequency 125–135 KH <inf xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">Z</inf> . The results show that the proposed class-E design could use the same values of C <inf xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">1</inf> , C <inf xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</inf> , L <inf xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">2</inf> for small and optimum load resistor with high efficient more than 80% All these three class-E power amplifiers will be presents in this work and will be simulate with the MATLAB/SIMULINK environment, and for further verification all the circuits will be simulated with electronics workbench MULTISIM 9.

References

7