A High Pressure Temperature-Jump Apparatus

A.D. Yu, Mario Waissbluth, R. A. Grieger

Review of Scientific Instruments · 1973 · 20 citations · 3 references

Concepts

Abstract

A technique is described for carrying out temperature-jump experiments at pressures up to 1000 atm, thus allowing the measurement of activation volumes for reactions in the μsec-sec range. The main features of the new technique are an electrical lead which has been tested to 2000 atm and 10 000 V ungrounded electrodes, and a flexible cell to transmit pressure from the external fluid to the solution.

References

3