Medical Physics · 2007 · 104 citations · 24 references
MedicineBiomedical EngineeringRadiologyDiagnostic ImagingDiagnostic RadiologyRadiation DetectionBreast ImagingImage Quality AssessmentMammography PhysicsX-ray DetectionDetective Quantum EfficiencyMammographic Image QualityMammographyModulation Transfer FunctionNoise Power Spectrum In MammographyPhoton CountingQuantum EfficiencySignal Collection
An objective analysis of image quality parameters was performed for six digital mammography systems. The presampled modulation transfer function (MTF), normalized noise power spectrum (NNPS), and detective quantum efficiency (DQE) for the systems were determined at different doses, for 28 kVp with a Mo/Mo or W/Al target/filter combination and 2 mm of additional aluminium filtration. The flat-panel units have higher MTF and DQE in the mid to high frequency range than standard CR systems. The highest DQE, over the whole dose range, is for the slit-scanning direct photon counting system. Dual-side read CR can overcome the inherent x-ray absorption and signal collection limitations of standard CR mammography, improving the low-frequency DQE by 40%, to the same level as full-field systems, but it does not improve the poor spatial resolution of phosphor.
24
Absorption and noise in x-ray phosphors
Robert K. Swank · Journal of Applied Physics · 1973 · 343 citations
Natural Sciences, Materials Science, Radiation Detection +16
DQE(<i>f</i>) of four generations of computed radiography acquisition devices
James T. Dobbins, David L. Ergun, Lois Rutz et al. · Medical Physics · 1995 · 307 citations