Analytical Chemistry · 1998 · 66 citations · 27 references
An amplified flow immunoassay (AFIA) was developed for cocaine, which combines a noncompetitive immunoenzymometric assay (IEMA) with an on-line detection of the enzyme label alkaline phosphatase (ALP) by a substrate-recycling biosensor. In the IEMA, the analyte cocaine first binds to a labeled polyclonal anti-cocaine antibody. Then, the excess labeled antibody is separated on an affinity column that contains a perfusion chromatography carrier modified by immobilized cocaine. The unbound complexes of the analyte cocaine with the ALP-labeled antibody are detected postcolumn. The detector senses phenol produced by ALP from phenyl phosphate. As detector, an amperometric substrate-recycling biosensor was used, which consists of a Clark-type oxygen electrode covered by tyrosinase and pyrroloquinoline quinone-dependent glucose dehydrogenase. The lower limit of detection is 380 pM (38 fmol) for cocaine. The sampling rate is 26/h. Cocaine could be detected from "real samples" with an imprecision of +/- 10% (n = 3) and with a recovery of 49 +/- 3% for various concentrations. AFIA is generally important as a new approach for the fast detection of picomolar concentrations of haptens.
27
Christian G. Bauer, Arkadi V. Eremenko, Eva Ehrentreich‐Förster et al. · Analytical Chemistry · 1996 · 193 citations
Engineering, Pesticide-residue Analysis, Bioelectrochemistry +22
Enzyme amplification for immunoassays
Axel Johannsson, D. Ellis, David L. Bates et al. · Journal of Immunological Methods · 1986 · 123 citations
Capillary Electrophoresis-Based Immunoassay for Cortisol in Serum
Dieter Schmalzing, Wassim Nashabeh, Xian-Wei. Yao et al. · Analytical Chemistry · 1995 · 105 citations