Publication | Open Access
Validity and test-retest reliability of manual goniometers for measuring passive hip range of motion in femoroacetabular impingement patients.
259
Citations
40
References
2010
Year
The present study suggests that goniometer-based assessments considerably overestimate hip joint ROM by measuring intersegmental angles (e.g., thigh flexion on trunk for hip flexion) rather than true hip ROM. It is likely that uncontrolled pelvic rotation and tilt due to difficulties in placing the goniometer properly and in performing the anatomically correct ROM contribute to the overrating of the arc of these motions. Nevertheless, conventional manual goniometers can be used with confidence for longitudinal assessments in the clinic.
| Year | Citations | |
|---|---|---|
Page 1