Journal of Liquid Chromatography · 1978 · 15 citations · 7 references
PharmacyPharmaceutical AnalysisPharmacologyAnesthesiaChromatographic AnalysisClinical PharmacologyAnesthetic Drug PharmacologyAnesthetic PharmacologyChromatographic Analysis Of PharmaceuticalsInternal Standard AnalysisIntravenous LidocaineReversed-phase ColumnsSensitivity
Abstract A sensitive and precise high-pressure liquid chromatographic method for the simultaneous quantitation of lidocaine (lignocaine) and thiopental in 0.5 ml samples of plasma is described. Extraction was conducted at pH 7.0 using ethyl acetate, and bupivicaine was employed as an internal standard. The drugs were eluted from a reversed-phase column using a mobile phase consisting of aceton-itrile/0.2 M phosphate buffer pH 4.0 (1:9) at a flow rate of 1.0 ml/min. The drugs were detected and quantified by their UV absorption at 205 nm. The sensitivity limits of detection for lidocaine and thiopental were 1 ug/ml and 5 ug/ml, respectively. Atropine, caffeine and meperidine were found to have interfering retention times.
7
The Pharmacological Basis of Therapeutics
American Journal of Health-System Pharmacy · 1976 · 8.2K citations
Procainamide dosage schedules, plasma concentrations, and clinical effects
Jan Koch‐Weser · JAMA · 1971 · 41 citations