Variation of zeta‐potential with temperature in fused‐silica capillaries used for capillary electrophoresis

Christopher J. Evenhuis, Rosanne M. Guijt, Philip J. Marriott, Paul R. Haddad

Electrophoresis · 2006 · 48 citations · 13 references

DOIFull text

Open access

Concepts

Abstract

The temperature variation of electroosmotic mobility corrected for the effects of Joule heating (muEOF0) was employed to investigate the variation of the zeta-potential (zeta) with temperature in fused-silica capillaries. Experimentally determined values for zeta increased at 0.39% per degrees C, a rate that is about four to five times smaller than reported previously. Experimentally determined values of zeta were directly proportional to the absolute temperature although values were also influenced slightly by changes to the dielectric constant. It was found that the effective charge density at the inner surface of the capillary was independent of temperature.

References

13