Medical Physics · 2005 · 49 citations · 36 references
MedicineRadiation OncologyRadiation DosimetryRadiation PhysicsTreatment PlanningElectron Beam PhysicsElectron DosimetryExternal Beam DosimetryPhoton DosimetrySmall Field DosimetryClinical Electron DosimetryDepth Dose DistributionDose Depth CurveElectron Beam SourcesElectron SourcesExponential Curve FittingPercentage Depth DoseSmall Field Electron Dosimetry
The purpose of the present study is to characterize electron contamination in photon beams in different clinical situations. Variations with field size, beam modifier (tray, shaping block) and source-surface distance (SSD) were studied. Percentage depth dose measurements with and without a purging magnet and replacing the air by helium were performed to identify the two electron sources that are clearly differentiated: air and treatment head. Previous analytical methods were used to fit the measured data, exploring the validity of these models. Electrons generated in the treatment head are more energetic and more important for larger field sizes, shorter SSD, and greater depths. This difference is much more noticeable for the 18 MV beam than for the 6 MV beam. If a tray is used as beam modifier, electron contamination increases, but the energy of these electrons is similar to that of electrons coming from the treatment head. Electron contamination could be fitted to a modified exponential curve. For machine modeling in a treatment planning system, setting SSD at 90 cm for input data could reduce errors for most isocentric treatments, because they will be delivered for SSD ranging from 80 to 100 cm. For very small field sizes, air-generated electrons must be considered independently, because of their different energetic spectrum and dosimetric influence.
36
A pencil beam model for photon dose calculation
Anders Ahnesjö, M. Saxner, Avo Trepp · Medical Physics · 1992 · 327 citations
A virtual photon energy fluence model for Monte Carlo dose calculation
Matthias Fippel, Freddy Haryanto, O. Dohm et al. · Medical Physics · 2003 · 196 citations
Christine L. Hartmann Siantar, R. S. Walling, T.P. Daly et al. · Medical Physics · 2001 · 169 citations · Full text