EngineeringNuclear EngineeringInstrumentationRadiation MeasurementRadiation ProtectionAlpha Particle DetectionDosimetry InstrumentationRadiation DetectorsRadiation Dose MeasurementRadiation Monitoring SystemsInstrument Performance CharacterizationInstrument Performance MechanismsIonizing Radiation DosimetryRadiation Detection InstrumentationRadon MeasurementRadon Monitoring
An instrument performance test has been carried out for a radon measuring system made in Hungary. The system measures radon using the alpha-track detection technique. It consists of three parts: the passive detector, the etching unit and the evaluation unit. A CR-39 detector is used as the radiation detector. Alpha-track reading and data analysis are carried out after chemical etching. The following subjects were examined in the present study: (1) radon sensitivity, (2) performance of etching and evaluation processes and (3) thoron sensitivity. The radon sensitivity of 6.9 x 10(-4) mm(-2) (Bq m(-3) d)(-1) was acceptable for practical application. The thoron sensitivity was estimated to be as low as 3.3 x 10(-5) mm(-2) (Bq m(-3) d)(-1) from the experimental study.