Improved determination of aluminum in serum and urine with use of a stabilized temperature platform furnace.
60
Citations
0
References
1982
Year
Stabilized Temperature PlatformHealthy SubjectsMedicineBioanalysisMass SpectrometryLaboratory MedicineChemical PathologyBiostatisticsAnalytical ChemistryMatrix InterferenceClinical ChemistryChronic Kidney DiseaseNephrologyChromatography
This method for determining aluminum in serum and urine is essentially free from matrix interference and gives a linear response with concentration to at least 500 micrograms/l. Use of a stabilized temperature platform (L'vov platform, Perkin-Elmer Corp.) to approach a "steady-state" temperature, addition of matrix modifiers [especially Mg(NO3)2], and the use of peak area integration all helped substantially diminish spectral interference. With the platform furnace, serum protein concentrations as great as 260 g/L did not interfere with the determination of Al. The within- and between-assay precision (CV) was less than or equal to 3.5% and less than or equal to 7.4%, respectively. Analytical recovery of Al added to serum ranged between 95 and 101% throughout the linear calibration range (to 500 micrograms/L), either when measured directly from the standard curve or by the method of standard additions. The reference interval for Al in 28 healthy subjects was 2-14 micrograms/L (mean 6.5, SD 4.1 micrograms/L), and for 130 patients on hemodialysis, 20-550 micrograms/L (mean 87.5, SD 62.5 micrograms/L).