Clinical Chemistry · 1983 · 10 citations · 0 references
MedicineBiochemistryBiostatisticsDiabetologyGood Laboratory PracticeMetabolic RegulationBioanalytical Method ValidationDiabetes MellitusGlycemic ControlGlycemic MonitoringMetabolic EvaluationMethod ComparisonBlood Glucose RegulationGlucose Analysis In Body FluidsGlucose Data InterpretationLaboratory Plasma Glucose TestingOrdinal ArithmeticOrdinal ClassificationOrdinal ComparisonStatistical Comparison Of Measurement Results
We report a new statistical tool for comparing several dry-reagent strip procedures for whole blood glucose, which produce data in both digital and ordinal form, with results by the well-studied hexokinase-glucose-6-phosphate dehydrogenase procedure coupled to NAD+-NADH. Our use of "ordinal comparison unit" allows for a more equitable comparison of such data. These strip procedures produce biases of -2.21 to 1.74 ordinal comparison units over the range of glucose values corresponding to hypoglycemia and hyperglycemia, as compared with results by the hexokinase procedure, but they are essentially equivalent when compared with each other.