Transcatheter access route options for treatment of degenerated mitral valve prosthesis with a balloon‐expandable biological valve

Ulrich Schäfer, Christian Frerker, Ralf Bader, Michael Schmoeckel, Cord Busse, Karl‐Heinz Kück

Catheterization and Cardiovascular Interventions · 2012 · 10 citations · 13 references

DOIFull text

Open access

Concepts

Abstract

OBJECTIVES: There is an ongoing discussion if valve-in-valve implantations into failing mitral xenografts should be performed only via a transapical approach. BACKGROUND: In selected high-risk patients, transcatheter valve implantation is an emerging therapy for degenerated mitral xenografts in need of repeated valve replacement. METHODS: Here we describe three different strategies to treat a failing mitral biological prosthesis (Carpentier Edwards S.A.V) by implantation of a SAPIEN-XT valve, which was performed in three different patients. As access site, a transjugular/transseptal approach, a transfemoral/transseptal approach, and a transapical approach were chosen. RESULTS: With all three approaches, valve-in-valve implantation was successfully achieved. The acute hemodynamic effects were excellent and according to VARC there was no adverse event recorded during a 6 months follow-up. CONCLUSIONS: Feasibility of all three approaches is demonstrated and might be taken into consideration for treatment in various patients.

References

13